首页> 外文会议>5th International Symposium on Test and Measurement (ISTM/2003) Vol.4 Jun 1-5, 2003 Shenzhen, China >A Calibration Theory and Method Named Nose-to-nose Based on the Agilent Broadband Sampling Oscilloscopes
【24h】

A Calibration Theory and Method Named Nose-to-nose Based on the Agilent Broadband Sampling Oscilloscopes

机译:基于安捷伦宽带采样示波器的标称从鼻子到鼻子的校准理论和方法

获取原文
获取原文并翻译 | 示例

摘要

The nose-to-nose (NTN) calibration theory is introduced by Ken Rush and Dr. Jan Verspecht in 1994 after several years study, it is an accurate method to determine the impulse response of broadband sampling oscilloscopes. The method is based on the hypothesis "sampler's kick-out pulse equals oscilloscope impulse response". We have studied the theory from stimulation to experiment. In this paper, the kick-out pulse and the response of the kick-out pulse are presented using PSPICE and MATLAB software. And two same Agilent 86100 types are connected to obtain the response waveform of the kick-out pulse. Although data processing is not perfect yet, simple introductions are given in this artical.
机译:Ken Rush和Jan Verspecht博士在经过数年的研究之后,于1994年提出了鼻-鼻(NTN)校准理论,这是一种确定宽带采样示波器脉冲响应的准确方法。该方法基于以下假设:“采样器的触发脉冲等于示波器的脉冲响应”。我们已经研究了从刺激到实验的理论。在本文中,使用PSPICE和MATLAB软件介绍了启动脉冲和启动脉冲的响应。并连接了两个相同的Agilent 86100型以获得反冲脉冲的响应波形。尽管数据处理还不完善,但本文中仅作了简单介绍。