The purpose of this research was to study on the design and construction of infusion pump calibrator. This research has adopted the principle of physics, electronics and microcontroller. The designed and constructed the infusion pump calibrator was composed of three main parts: 1)the control part consisting of micro switches for selected the operating mode, 2) signals detection part composing of load cells, pressure sensor for measuring flow rate and occlusion alarm respectively and3) the processing and display part comprising of microcontroller AVR ATmega1280 with C Language program and LCD sequentially. The results of functional testing were compared with standard infusion devices analyzer Metron Lagu showed that the average percentage error of flow rate measuring and occlusion pressure tests were 0.45 percent and 0.21 percent successively.
机译:这项研究的目的是研究输液泵校准器的设计和构造。这项研究采用了物理学,电子学和微控制器的原理。设计和制造的输液泵校准器由三个主要部分组成:1)控制部分由用于选择操作模式的微动开关组成; 2)组成称重传感器的信号检测部分,分别用于测量流量的压力传感器和闭塞报警器3)处理和显示部分,其包括依次具有C语言程序的微控制器AVR ATmega1280和LCD。将功能测试的结果与标准输液设备分析仪Metron Lagu进行比较,结果表明,流量测量和阻塞压力测试的平均百分比误差分别为0.45%和0.21%。