首页> 外文会议>4th International symposium on advanced topics in electrical engineering (ATEE 2004) >VERIFICAREA CORECTITUDINII M?SUR?RILOR EFECTUATE ?NLABORATOARELE DE METROLOGIE
【24h】

VERIFICAREA CORECTITUDINII M?SUR?RILOR EFECTUATE ?NLABORATOARELE DE METROLOGIE

机译:计量实验室中进行的测量准确性的验证

获取原文
获取原文并翻译 | 示例

摘要

This paper presents the manner to demonstrate the competence of the metrological laboratoriesrninvolved in the dissemination of the DC voltage unit. The measurements were made for the values ofrn1,018 V and 10 V. It was calculated the standard deviation for the measured voltages of eachrnlaboratory related to the reference voltage and were determined the values of the indexes for thesernlaboratories. It was estimated the uncertainty for the standard deviation. In the paper are shown thernresults of the measurements for each of the six laboratories that were participated and for thernreference laboratory. As itinerant standard was used an electronic DC voltage calibrator with therncapability to generate two output voltages: 1,018 V and respectively 10 V. The result of the analysis ofrnthe results that were sent by these laboratories demonstrate that all laboratories were inside of thernacceptance limits for the value of the index. Thus, all the laboratories are capable to transmitrncorrectly the DC voltage unit, inside of the uncertainty assumed by each laboratory. All thernmeasurements made by the participant laboratories are compatible.
机译:本文提出了一种方法来证明参与直流电压单位的传播的计量实验室的能力。分别对rn1,018 V和10 V的值进行了测量。计算了与参考电压相关的每个实验室的测量电压的标准偏差,并确定了实验室的指标值。估计了标准偏差的不确定性。本文显示了所参与的六个实验室和参考实验室的测量结果。作为巡回标准,使用了能够产生两个输出电压的电子直流电压校准器:1,018 V和分别为10V。这些实验室发送的结果的分析结果表明,所有实验室均在该值的可接受范围内索引。因此,所有实验室都能够在每个实验室承担的不确定性范围内正确传输直流电压单位。参与实验室所做的所有测量都是兼容的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号