One out of the six fixtures installed showed a deviation of 7°from the planned axis direction, of 6 mm. from the planned maxillar entry and of 8 mm. from the planned zygomatic exit. Considering the other five fixtures, the worst(best) case values were 3.1(0.6)°? 2.1(0.7) mm. and 2.7(0.8) mm. respectively. These results indicate that acceptable accuracy is to be expected from zygoma fixture drilling templates.
机译:安装的六个固定装置中有一个显示出与计划的轴方向的7°偏差为6 mm。从计划的上颌骨入口和8毫米。从计划的骨出口退出。考虑其他五个夹具,最坏(最佳)情况值为3.1(0.6)°? 2.1(0.7)毫米和2.7(0.8)毫米。分别。这些结果表明,从zygoma夹具钻孔模板可以预期达到可接受的精度。