IR radiometers are used for the measurement of IR intensity in military, industry, academic and research domains. For real-time, noncontact, accurate measurement with high sensitivity SR 5000 radiometer is employed. It covers the waveband from 0.4-14.5 μm and has a provision to accommodate many filters and detectors. The measurements are carried out with respect to internal black body, maintained at room temperature. As any measurement equipment requires repeatable, reproducible & accurate output, calibration at regular interval is a mandatory requirement. In the current paper the calibration of radiometer in radiometric mode using different filter bands at a distance of 80 and 3 meters is described. The calibration is done at different temperatures and Planck's function is calculated. The measured results are validated with the Planck's function. This paper also describes the convergence of results for 80 and 3 meters.
机译:红外辐射计用于军事,工业,学术和研究领域的红外强度测量。为了实现实时,非接触式,高灵敏度SR 5000辐射计的精确测量。它覆盖了0.4-14.5μm的波段,并具有容纳许多滤波器和检测器的功能。相对于保持在室温下的内部黑体进行测量。由于任何测量设备都需要可重复,可再现和准确的输出,因此必须定期进行校准。在当前的论文中,描述了在80米和3米距离处使用不同的滤带在辐射模式下对辐射计进行的校准。在不同的温度下进行校准,并计算出普朗克函数。测量结果通过普朗克函数进行验证。本文还描述了80米和3米结果的收敛性。