This paper presents a method for the control of the variable burst biphasic pulse of a bladder stimulator. The stimulator is used to pass current through the tissue and to generate useful action potentials. The binary-weighted digital-to-analog converter combined with a current mirror has been employed as a microstimulator because of its higher linearity without requiring the decoding of digital inputs. Given that the use of a biphasic pulse could prevent ion-charge accumulation in tissues, two pairs of switches controlled by different clock phases are implemented to provide the biphasic electrical stimulation pulses. According to measurement results, the pulse frequency can be programmed between 1.49 Hz and 47.66 Hz, the burst frequency can be controlled from 190.8 Hz to 763 Hz, and the pulse width can be adjusted between 21 µs and 325 µs. These stimulation parameters are adapted by the clock divider and by the number of controlled bits in the digital circuits.
机译:本文提出了一种控制膀胱刺激器可变脉冲双相脉冲的方法。刺激器用于使电流流过组织并产生有用的动作电位。二进制加权数模转换器与电流镜相结合已被用作微刺激器,因为它具有更高的线性度,而无需对数字输入进行解码。考虑到使用双相脉冲可以防止组织中离子电荷的积累,实施了由不同时钟相位控制的两对开关以提供双相电刺激脉冲。根据测量结果,可以将脉冲频率编程为1.49 Hz至47.66 Hz,可以将突发频率控制在190.8 Hz至763 Hz,并且可以将脉冲宽度调整为21 µs至325 µs。这些刺激参数由时钟分频器和数字电路中受控位的数量进行调整。