首页> 外文会议>2012 42nd European Microwave Conference. >Wireless power transfer for a miniature gastrostimulator
【24h】

Wireless power transfer for a miniature gastrostimulator

机译:微型胃刺激器的无线电源传输

获取原文
获取原文并翻译 | 示例

摘要

In this work, wireless power transfer issues were investigated for a gastroparesis management system consisting of an endoscopically-implantable miniature batteryless/rechargeable wireless gastrostimulator and a wearable battery-operated transmitter module. Output power for stimulation and input power consumption in the transmitter as well as the wireless power transfer efficiency of two coil antennas with significantly different sizes were examined. With an implant antenna coil of 1×3.5 cm2 and 5-cm radius transmitter antenna coils of varying coil turns, power transfers and efficiencies at the individually-tuned resonance of 1.3MHz were experimentally obtained at various coil separation distances from 4 to 10 cm between two coils. Taking power consumption in the wearable module and tolerance of parasitic capacitance in the environment into consideration, a 17-turn transmitter coil was chosen. RF losses through tissues were also examined with saline mimicking the abdomen tissues. At a distance of 10 cm as the worst scenario, a load current 3.02 mA was achieved which was still sufficient for gastrostimulation. Measurements of output voltage and efficiency distribution were mapped in 3-D to examine the effects of antenna misalignment due to body and stomach motion. This work provided insights into a complete system design for an endoscopically-implantable gastrostimulator that could relieve the suffering of many gastroparesis patients.
机译:在这项工作中,研究了胃轻瘫管理系统的无线功率传输问题,该系统由内窥镜植入式微型无电池/可充电无线胃刺激器和可穿戴电池供电的发射器模块组成。检查了发射器中用于刺激的输出功率和输入功率消耗,以及尺寸明显不同的两个线圈天线的无线功率传输效率。通过使用1×3.5 cm 2 的植入天线线圈和半径为5cm的线圈匝数变化的发射器天线线圈,分别在各种线圈上实验获得了1.3MHz的单调谐谐振下的功率传输和效率两个线圈之间的分离距离为4到10厘米。考虑到可穿戴模块的功耗以及环境中的寄生电容容差,选择了17匝发射线圈。还通过模仿腹部组织的生理盐水检查了通过组织的射频损失。最坏的情况是在10 cm的距离上达到3.02 mA的负载电流,这仍然足以进行胃刺激。将输出电压和效率分布的测量结果映射到3-D中,以检查由于身体和胃部运动引起的天线未对准的影响。这项工作为可减轻许多胃轻瘫患者痛苦的内窥镜植入式胃刺激器的完整系统设计提供了见识。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号