Abstract: A new method has been developed to calibrate tristimulus colorimeters for high accuracy color measurements. Instead of traditional lamp standards, modern, high accuracy detector standards are suggested for calibration. After high accuracy absolute spectral response determination of the tristimulus receivers, color (spectral) correction and peak (amplitude) normalization can minimize uncertainties caused by imperfect realizations of the Commission Internationale de l'Eclairage color matching functions. As a result of the corrections, stable light sources of different spectral power distributions can be measured with an accuracy dominated by the sub 0.1% uncertainty of novel spectral response determinations.!1
机译:摘要:已开发出一种用于校准三色度色度计以进行高精度色彩测量的新方法。代替传统的灯泡标准,建议使用现代的高精度检测器标准进行校准。在确定了三刺激接收器的高精度绝对光谱响应之后,色彩(光谱)校正和峰值(振幅)归一化可以最大程度地减少由国际照明委员会色彩匹配功能的不完善实现所引起的不确定性。校正的结果是,可以以新光谱响应确定的亚0.1%不确定度以下的精度来测量具有不同光谱功率分布的稳定光源!1