Methods of verifying the output of the treatment planning system used for high dose rate (HDR) prostate brachytherapy

机译:验证用于高剂量率(HDR)前列腺近距离放射治疗的治疗计划系统输出的方法

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: To investigate the relationships between the planning target volume (PTV) and the total reference air kerma (TRAK) required to irradiate the prostate using the Nucletron microSelection 192Ir radioactive stepping source afterloader for the treatment of prostate cancer. Methods: Using 54 patient plans, the relationship between the treatment planning system (TPS) TRAK (TRAK TPS) and the estimate and planimetry prostate volumes was investigated. Linear regression analysis was used to predict the TRAK (TRAK pred) using the prostate gland height, width and length (HWL) measurements. The TRAK was corrected using a conformal index (COIN) of the plan to improve the accuracy of the model. A further 54 patient plans were then used to evaluate the outcome of the linear regression lines. Results: A formula was proposed (HxWxLxπ/5.19) to estimate the prostate volume, which showed a mean deviation from the planimetry volume of 0.0139 ± 3.339 cm 3, where 69% was within ±10% of planimetry, and 96% within ±20%. Scatter graphs of TRAK TPS for 54 plans, showed a positive linear correlation, r = 0.98, using the planimetry volume, and r = 0.95 for the estimated volume. The further 54 treatment plans used for evaluation showed the TRAK pred was accurately predicted to ±10% for 100% of plans using the planimetry volume and 81% using the prostate volume estimate. Conclusions: The verification method proposed was found to be accurate and independent of TPS parameters, with suitable acceptance margins of ±10%.
机译:目的:研究使用Nucletron microSelection 192Ir放射性步进源后装器治疗前列腺癌时,计划目标体积(PTV)与辐射前列腺所需的总参考空气比释动能(TRAK)之间的关系。方法:使用54个患者计划,研究了治疗计划系统(TPS)TRAK(TRAK TPS)与估计量和平面测量前列腺体积之间的关系。线性回归分析用于通过前列腺高,宽和长(HWL)测量来预测TRAK(TRAK pred)。使用计划的保形索引(COIN)纠正了TRAK,以提高模型的准确性。然后,使用另外的54个患者计划来评估线性回归线的结果。结果:提出了一个公式(HxWxLxπ/ 5.19)来估计前列腺体积,该公式显示出与体积的平均偏差为0.0139±3.339 cm 3,其中69%的体积在平面的±10%之内,而96%的体积在±20%的范围内%。 TRAK TPS的54个平面图的散点图显示正线性相关,使用平面规划的体积,r = 0.98,对于估计的体积,r = 0.95。用于评估的另外54个治疗计划显示,使用平面测量术体积将100%计划准确预测为TRAK pred为±10%,使用前列腺体积估算将其准确预测为81%。结论:所提出的验证方法被发现是准确的,并且与TPS参数无关,具有±10%的适当接受余量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号