Fractionated high dose rate afterloading brachytherapy for prostate cancer requires a robust means of catheter fixation with good quality assurance. Catheter position and dosimetry has been formally evaluated in 20 consecutive patients representing a total of 332 catheters undergoing two HDR afterloading brachytherapy fractions over 36 h. The mean interfraction movement of catheters as measured by external length was less than 1 mm, but within the prostate on consecutive CT scans there was a mean interfraction movement of 11.5 mm away from the prostate base. This has a significant impact on implant dosimetry as measured by D90 and the COIN index, unless corrected by repositioning the catheters.
机译:前列腺癌分装近距离放疗后的高剂量率分级需要一种可靠的导管固定方法,并具有良好的质量保证。导管位置和剂量测定已在20位连续患者中进行了正式评估,代表总共332条导管在36小时内接受两次HDR后近程放射治疗。通过外部长度测量的导管的平均屈光运动小于1 mm,但是在连续的CT扫描中,在前列腺内部,距前列腺根的平均屈光运动为11.5 mm。除非通过重新放置导管进行校正,否则这对通过D90和COIN指数测量的植入物剂量有重大影响。