首页> 外文期刊>Applied optics >Study of filtered light on potential retinal photic hazards with operation microscopes used for ocular surgery
【24h】

Study of filtered light on potential retinal photic hazards with operation microscopes used for ocular surgery

机译:眼科手术用手术显微镜对滤过光对视网膜潜在光害的研究

获取原文
获取原文并翻译 | 示例
           

摘要

There have been numerous reports of retinal photic injury from operation microscopes used during cataract surgery. The risk of injury has been associated with the intensity of the light directed into the eye, short-wavelength emission, user technique, exposure time, and direct axial lighting. We evaluated use of light transmission filters to modify a tungsten-halogen light source spectrum to reduce the risk of retinal photic injury. A two-light source filter combination was found with a color profile acceptable for intraocular surgery that reduces the risk of retinal photic injury by a factor of approximately 2.5. [References: 19]
机译:白内障手术期间使用手术显微镜已经有许多视网膜光损伤的报道。受伤的风险与进入眼睛的光强度,短波发射,用户技术,曝光时间和直接轴向照明有关。我们评估了使用光透射滤光片来修改卤素钨光源光谱以降低视网膜光性损伤的风险。发现了一种双光源滤光片组合,其色彩特征可用于眼内手术,可将视网膜光性损伤的风险降低约2.5倍。 [参考:19]