Usefulness of the virtual monochromatic image in dual-energy spectral CT for decreasing renal cyst pseudoenhancement: A phantom study

机译:双能谱CT中虚拟单色图像对减少肾囊肿假增强的有用性:一项幻像研究

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE. The purpose of this study was to determine the effect that dual-energy spectral CT has on renal cyst pseudoenhancement with a renal cyst phantom. MATERIALS AND METHODS. Two renal compartments within a CT phantom were filled with 40- and 240-HU solutions of diluted contrast material, mimicking the nonenhanced phase and nephrographic phase of MDCT, respectively. A saline-filled tapering cylindric cyst model was used, simulating cysts of varying diameters (range, 0.7-3.0 cm) suspended in the renal compartments. The phantom was scanned using dual-energy CT (DECT) (80 and 140 kVp) and conventional, single-energy CT (120 kVp) with the same 64-MDCT scanner. Attenuation increase was measured for a cyst model within 40- and 240-HU dilutions. We compared the attenuation increase obtained on virtual monochromatic images from DECT with conventional, single-energy MDCT images using a paired Student t test. RESULTS. Pseudoenhancement occurred in the conventional single-energy CT acquisitions, with water attenuation increase of 42.44 ± 4.03 HU measured at 120 kVp. In virtual monochromatic images of DECT, we observed less pseudoenhancement with water attenuation increase of 21.51 ± 6.18 HU at 70 keV. Both acquisitions yielded a p value less than 0.0001. CONCLUSION. We achieved less pseudoenhancement of renal cysts with virtual monochromatic images of DECT compared with conventional single-energy CT. This method may be useful when an accurate differentiation between small renal cysts and solid masses is critical.
机译:目的。这项研究的目的是确定双能谱CT对带有肾囊肿体模的肾囊肿假性增强的作用。材料和方法。 CT体模中的两个肾区室分别充满40和240 HU稀释的对比剂溶液,分别模拟MDCT的未增强期和肾造影相。使用了盐水填充的锥形圆柱囊肿模型,模拟了直径不等(0.7-3.0 cm)的囊肿悬浮在肾腔中。使用双能CT(DECT)(80和140 kVp)和传统的单能CT(120 kVp)并使用相同的64-MDCT扫描仪扫描体模。对于40-HU和240-HU稀释度的囊肿模型,测量了衰减增加。我们使用配对的Student t检验比较了DECT的虚拟单色图像与常规的单能量MDCT图像获得的衰减增加。结果。在常规单能CT采集中出现了伪增强,在120 kVp下测得的水衰减增加了42.44±4.03 HU。在DECT的虚拟单色图像中,我们在70 keV处观察到较少的伪增强,水衰减增加了21.51±6.18 HU。两次收购均产生小于0.0001的p值。结论。与传统的单能CT相比,通过DECT的虚拟单色图像,我们获得的肾囊肿假性增强更少。当小肾囊肿和实性肿块之间的准确区分非常关键时,此方法可能有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号