首页> 外文期刊>European journal of clinical nutrition >Evaluation of the amount of residual lipid emulsion in chambers of flushed totally implantable venous access devices using fluorescence imaging
【24h】

Evaluation of the amount of residual lipid emulsion in chambers of flushed totally implantable venous access devices using fluorescence imaging

机译:使用荧光成像评估冲洗完全可植入静脉接入装置的腔室中的残留脂质乳液量

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of the present study was to use a quantitative fluorescence imaging technique to evaluate the invisible amount of residual lipid emulsion in port chambers flushed with various fundamental protocols. Chambers were filled with lipid emulsion containing indocyanine green and then flushed with 5-70 mL of normal saline. Chambers were flushed at various speeds (15-60 mL/min), with a time interval of 1 or 3 s between boluses, and with varying directions of flow. The slower the flushing speed, the more lipid emulsion that remained. Pulsatile flushing with either time interval did not decrease the residual amounts, and the areas well-cleansed after flushing were oriented to the bevel-opening direction. These findings suggest that to reduce the residual amount of lipid emulsion poured in a chamber, fast and furious flushing under continuous as opposed to pulsatile flushing is of paramount importance.
机译:本研究的目的是使用定量荧光成像技术来评价与各种基本方案冲洗的端口室中的残留脂质乳液的无形量。 腔室填充含有吲哚菁绿的脂质乳液,然后用5-70ml的甘盐冲洗。 腔室以各种速度(15-60ml / min)冲洗,在钢管之间的时间间隔为1或3秒,并且具有不同的流动方向。 冲洗速度较慢,仍然存在脂质乳液。 脉动用任一时间间隔冲洗未降低剩余量,并且冲洗后清洁的区域定向到斜面方向。 这些发现表明,减少倒入室内的脂质乳液的残留量,与脉动冲洗相反,连续冲洗的快速和激烈冲洗是至关重要的。