Deep brain stimulation lead fixation: a comparative study of the Navigus and Medtronic burr hole fixation device.

机译:深部脑刺激导线固定:Navigus和Medtronic毛刺孔固定装置的比较研究。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES:: To determine the extent of lead movement based on the type of burr hole fixation device used to secure the lead (Image-Guided Neurologics [IGN], Navigus versus Medtronic [Model 7495-51]). A randomized, blinded design of lead movement measurement was used. METHODS:: A clinical series of 58 individuals undergoing placement of a deep brain stimulation (DBS) system with a total of 71 operative sides were examined. Lead position was compared between 71 paired, sagital, digitized X-rays of lead position immediately before and after the lead was secured to the basecap. Lead movement was measured in a randomized, blinded fashion using the Siemens measurement tool at an 8x magnification rate. The presurgical planned target centered at the cross hatch of the reticules on a lateral X-ray served as the reference point to determine lead movement. RESULTS:: The overall mean lead movement was significantly less using the IGN (1.9mm), as compared to the Medtronic (3.3mm), fixation device. Moreover, the pattern of lead movement was significantly different between the two devices. That is, the majority of measured movements using IGN device was in the superior direction (74%), whereas the opposite was true for the Medtronic device (i.e., 62% with inferior movement). CONCLUSION:: The IGN burr hole fixation device is associated with significantly less movement when securing the lead. Probable mechanisms of movement are discussed.
机译:目的:根据用于固定导线的毛刺孔固定装置的类型确定导线移动的程度(图像引导神经系统[IGN],Navigus与Medtronic [Model 7495-51])。使用了随机,盲法的引线移动测量设计。方法:对58例接受深部脑刺激(DBS)系统植入的患者进行了临床研究,共手术侧71例。在将导线固定到基帽之前和之后,在71个成对的,矢状,数字化的X射线位置上比较了导线位置。使用西门子测量工具以8倍的放大倍率以随机,盲目方式测量引线移动。以横向X射线上的十字线的阴影线为中心的术前计划目标用作确定导线移动的参考点。结果:与Medtronic(3.3mm)固定装置相比,使用IGN(1.9mm)的整体平均导线移动明显更少。此外,两个设备之间的引线移动方式明显不同。也就是说,使用IGN设备测得的大多数运动是朝上方向(74%),而对于Medtronic设备,情况恰好相反(即,对于下等运动为62%)。结论:IGN毛刺孔固定装置与固定引线时的移动明显减少有关。讨论了可能的运动机理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号