首页> 外文期刊>BMC Surgery >Lead fixation in deep brain stimulation: comparison of three lead anchoring devices in China
【24h】

Lead fixation in deep brain stimulation: comparison of three lead anchoring devices in China

机译:深度脑刺激中的铅固定:中国三种铅锚固装置的比较

获取原文
           

摘要

The accuracy of deep brain stimulation (DBS) depends on precise electrode positioning, which has been pursued for ideal treatment outcomes. As a critical component of DBS, the fixation performance of lead anchoring devices has been widely studied. Possible reasons for lead shift were analyzed in the current study and we further provided effective solutions to reduce potential manual errors. Seventy-nine patients who received DBS implantations at the Ruijin Hospital from April to November 2017 were retrospectively reviewed. Intraoperative lead shifts were measured by C-arm fluoroscopy. Lead adjustment counts were recorded and compared among three lead fixation devices: Stimloc? (Medtronic, Minneapolis, MN, USA), TouchLoc (SceneRay, Suzhou, China), and the traditional lead anchoring device. Mean (± SD) distances of lead shifts were 0.29?±?2.42?mm in Stimloc devices, 0.43?±?0.55?mm in TouchLoc devices, and 1.52?±?1.05?mm in traditional devices (p??0.0001). Average numbers of adjustments in this series were 0.3?±?0.5 in Stimloc devices, 0.3?±?1.3 in TouchLoc devices, and 1.1?±?1.0 in traditional devices (p?=?0.0001). Pairwise comparisons among the three devices (TouchLoc vs. Stimloc: p?=?0.273; TouchLoc vs. Traditional: p?=?0.0001; Stimloc vs. traditional: p??0.0001) suggested significant differences, which were mainly attributed to the traditional devices. Three lead anchoring devices have been compared for their performance in the accuracy of lead fixation, in which the newly designed lead fixation devices have presented its advantages to the traditional one. In addition to the application of the Stimloc and TouchLoc devices, verification by C-arm fluoroscopy should be performed to provide an intuitive view of the depth deviation of electrode position during DBS electrode implantation.
机译:深部脑刺激(DBS)的准确性取决于电极的精确定位,这一直是理想的治疗结果所追求的。作为DBS的重要组成部分,铅锚固装置的固定性能已得到广泛研究。在当前的研究中分析了导致引线偏移的可能原因,我们进一步提供了有效的解决方案,以减少潜在的人为错误。回顾性分析了2017年4月至2017年11月在瑞金医院接受DBS植入术的79例患者。术中铅位移通过C型臂透视检查。记录并记录了三种导线固定装置之间的导线调整计数:Stimloc? (美国明尼苏达州明尼阿波利斯的美敦力公司),TouchLoc(中国苏州的SceneRay)以及传统的铅锚固装置。在Stimloc设备中,导线偏移的平均(±SD)距离为0.29?±?2.42?mm,在TouchLoc设备中为0.43?±?0.55?mm,在传统设备中为1.52?±?1.05?mm(p?<?0.0001) 。在该系列中,平均调整数在Stimloc设备中为0.3±±0.5,在TouchLoc设备中为0.3±±1.3,在传统设备中为1.1±±1.0(p = 0.0001)。三种设备之间的成对比较(TouchLoc与Stimloc:p≥0.273; TouchLoc与传统:p≥0.0001; Stimloc与传统:p≤0.0001)表明存在显着差异,这主要归因于传统设备。比较了三种铅锚固装置在铅固定精度方面的性能,其中新设计的铅锚固装置向传统铅锚固装置展现了其优势。除了应用Stimloc和TouchLoc设备外,还应通过C型臂透视检查进行验证,以提供DBS电极植入过程中电极位置深度偏差的直观视图。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号