首页> 外文期刊>Review of Scientific Instruments >Retrofitting an atomic force microscope with photothermal excitation for a clean cantilever response in low Q environments
【24h】

Retrofitting an atomic force microscope with photothermal excitation for a clean cantilever response in low Q environments

机译:在低Q环境中对原子力显微镜进行光热激发改造以实现干净的悬臂响应

获取原文
获取原文并翻译 | 示例
           

摘要

It is well known that the low-Q regime in dynamic atomic force microscopy is afflicted by instrumental artifacts (known as “the forest of peaks”) caused by piezoacoustic excitation of the cantilever. In this article, we unveil additional issues associated with piezoacoustic excitation that become apparent and problematic at low Q values. We present the design of a photothermal excitation system that resolves these issues, and demonstrate its performance on force spectroscopy at the interface of gold and an ionic liquid with an overdamped cantilever (Q < 0.5). Finally, challenges in the interpretation of low-Q dynamic AFM measurements are discussed.
机译:众所周知,动态原子力显微镜中的低Q范围受悬臂的压电声激发引起的仪器伪像(称为“峰林”)的困扰。在本文中,我们揭示了与压电声激励相关的其他问题,这些问题在低Q值下变得明显而成问题。我们提出了解决这些问题的光热激发系统的设计,并展示了其在金和离子液体与超阻尼悬臂(Q <0.5)的界面上的力谱性能。最后,讨论了低Q动态AFM测量的解释难题。