Automated microdilution MIC results, obtained with the Autoreader (Sensititre, Inc., Salem, N.H.) following 5 h of incubation, were compared with manually read, concurrent control MICs following 18 h of incubation in a three-laboratory comparative study. A total of 704 members of the family Enterobacteriaceae or similar gram-negative organisms were tested against 17 antimicrobial agents. Autoreader MICs were within 1 doubling dilution of control values in 92.9% of instances. Discrepancies of +/- 2 doubling dilutions and +/- 3 or greater doubling dilutions were noted in 4.5 and 2.6%, respectively, of the 7,687 drug-organism combinations analyzed. The majority of errors occurred when beta-lactam antimicrobial agents were tested with a variety of different species. MICs at 5 h, when Pseudomonas aeruginosa was used, were possible in only half the isolates tested and yielded data on only a limited number of drugs in the remaining instances. Excluding results obtained with penicillin and ampicillin, which were uniformly poor, Staphylococcus aureus Autoreader values were within +/- 1 doubling dilution of control values in 93.6% of instances, 5.4% varied by +/- 2 dilutions, and only 1% of test values by +/- 3 or more dilutions from control values for 82 isolates tested against nine antimicrobial agents. Of eight additional S. aureus isolates tested that were resistant to methicillin, only one was read correctly by the Autoreader, with results on the remaining seven appearing as either insufficient growth or as total resistance to all drugs tested. Interlaboratory reproducibility was excellent for selected isolates of S. aureus and gram-negative bacilli. The accuracy of the Sensititre Autoreader MIC results was comparable to that of other same-day quantitative systems for members of the family Enterobacteriaceae and S. aureus, while the economic and procedural advantages of the broth microdilution method was maintained.
机译:在三个实验室的比较研究中,将孵育5小时后使用自动读数器(Sensititre,Inc.,Salem,N.H.)获得的自动微量稀释MIC结果与人工阅读的并发对照MIC进行了比较。总共对704种肠杆菌科或类似革兰氏阴性生物的成员进行了针对17种抗菌剂的测试。在92.9%的情况下,自动阅读器的MIC在对照值的1倍稀释范围内。在分析的7,687种药物-有机物组合中,分别有4.5和2.6%的+/- 2倍稀释倍数和+/- 3倍或更大倍倍稀释倍数的差异。大多数错误是在使用各种不同物种测试β-内酰胺类抗菌剂时发生的。当使用铜绿假单胞菌时,在5小时内的MIC仅在测试的一半分离物中是可能的,而在其余情况下仅产生有限数量药物的数据。不包括使用均较差的青霉素和氨苄西林获得的结果,金黄色葡萄球菌自动读数器值在93.6%的情况下为对照值的+/- 1倍稀释范围内,5.4%的+/- 2稀释度变化范围内,仅占测试的1%相对于九种抗微生物剂测试的82个分离物,相对于对照值的+/- 3倍或更多稀释倍数的最高值。在另外八种对甲氧西林有抗药性的金黄色葡萄球菌分离株中,自动阅读器只能正确读取其中一株,其余的七种结果表现为生长不足或对所有被测药物的总耐药性。对于金黄色葡萄球菌和革兰氏阴性杆菌的选定分离株,实验室间的再现性极好。对于肠杆菌科和金黄色葡萄球菌,Sensititre Autoreader MIC结果的准确性可与其他当天定量系统的准确性相提并论,同时保持了肉汤微稀释法的经济和程序优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号