A linear diode array (JFD-5) for match line in vivo dosimetry in photon and electron beams; evaluation for a chest wall irradiation technique.

机译:用于在光子和电子束中进行体内剂量匹配的线性二极管阵列(JFD-5);胸壁照射技术的评估。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: In vivo dosimetry using thermoluminiscence detectors (TLD) is routinely performed in our institution to determine dose inhomogeneities in the match line region during chest wall irradiation. However, TLDs have some drawbacks: online in vivo dosimetry cannot be performed; generally, doses delivered by the contributing fields are not measured separately; measurement analysis is time consuming. To overcome these problems, the Joined Field Detector (JFD-5), a detector for match line in vivo dosimetry based on diodes, has been developed. This detector and its characteristics are presented. MATERIALS AND METHODS: The JFD-5 is a linear array of 5 p-type diodes. The middle three diodes, used to measure the dose in the match line region, are positioned at 5-mm intervals. The outer two diodes, positioned at 3-cm distance from the central diode, are used to measure the dose in the two contributing fields. For three JFD-5 detectors, calibration factors for different energies, and sensitivity correction factors for non-standard field sizes, patient skin temperature, and oblique incidence have been determined. The accuracy of penumbra and match line dose measurements has been determined in phantom studies and in vivo. RESULTS: Calibration factors differ significantly between diodes and between photon and electron beams. However, conversion factors between energies can be applied. The correction factor for temperature is 0.35%/ degrees C, and for oblique incidence 2% at maximum. The penumbra measured with the JFD-5 agrees well with film and linear diode array measurements. JFD-5 in vivo match line dosimetry reproducibility was 2.0% (1 SD) while the agreement with TLD was 0.999+/-0.023 (1 SD). CONCLUSION: The JFD-5 can be used for accurate, reproducible, and fast on-line match line in vivo dosimetry.
机译:背景与目的:在我们的机构中​​通常使用热发光检测器(TLD)进行体内剂量测定,以确定在胸壁照射过程中匹配线区域的剂量不均匀性。但是,TLD有一些缺点:无法进行在线体内剂量测定;通常,贡献场传递的剂量不会单独测量;测量分析非常耗时。为了克服这些问题,已经开发了联合场检测器(JFD-5),一种基于二极管的用于体内匹配线剂量测定的检测器。介绍了该检测器及其特性。材料与方法:JFD-5是5个p型二极管的线性阵列。用来测量匹配线区域中剂量的中间三个二极管以5mm的间隔放置。外部两个二极管的位置与中心二极管相距3 cm,用于测量两个贡献场中的剂量。对于三个JFD-5检测器,已经确定了不同能量的校准因子,以及非标准视野大小,患者皮肤温度和倾斜入射的灵敏度校正因子。半影和匹配线剂量测量的准确性已在幻像研究和体内确定。结果:二极管之间以及光子束和电子束之间的校准因子差异很大。但是,可以应用能量之间的转换因子。温度的校正系数为0.35%/摄氏度,斜入射的校正系数最大为2%。用JFD-5测量的半影与薄膜和线性二极管阵列的测量非常吻合。 JFD-5体内匹配线剂量学重现性为2.0%(1 SD),而与TLD的一致性为0.999 +/- 0.023(1 SD)。结论:JFD-5可用于准确,可重现和快速的在线匹配线体内剂量测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号