Spatial resolution of a stacked radiochromic film dosimeter.

机译:堆叠式放射色膜剂量计的空间分辨率。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The spatial resolution of stacked radiochromic film dosimeters, which have increased sensitivity from a single layer radiochromic film detector, has been studied. METHODS: A 5-layer film which can easily be constructed provided a 4.3-times increase in sensitivity over a single layer film at 670 nm readout wavelength which meant that doses as low as 0.6 Gy could be measured with an accuracy of +/-4% with the stacked dosimeter. The spatial resolution was tested by comparison of the 80%/20% penumbral widths of a 5 x 5 cm 6 MV X-ray field. RESULTS: The MD-55-2 film measured the penumbral width as 3.0 mm whereas the 5-layer stack dosimeter measured the same penumbra as 3.2 mm.Conclusion: The stack dosimeter can provide useful in vivo information such as the position of a diverging beam edge for treatments around critical structures such as eyes during the first fraction of treatment.
机译:目的:研究了单层放射性变色膜检测器提高了灵敏度的堆叠式放射性变色膜剂量计的空间分辨率。方法:可以轻松构建的5层膜在670 nm读出波长下的灵敏度比单层膜提高4.3倍,这意味着可以以+/- 4的精度测量低至0.6 Gy的剂量%与堆积的剂量计。通过比较5 x 5 cm 6 MV X射线场的80%/ 20%半影宽度测试了空间分辨率。结果:MD-55-2膜的半影宽为3.0 mm,而5层堆叠剂量计的半影为3.2 mm。结论:堆叠剂量计可提供有用的体内信息,例如发散光束的位置在治疗的第一阶段中,在边缘等关键结构周围进行治疗的边缘。