PTV margins should not be used to compensate for uncertainties in 3D image guided intracavitary brachytherapy.

机译:在3D图像引导的腔内近距离放射治疗中,不应使用PTV边缘来补偿不确定性。

获取原文
获取原文并翻译 | 示例
           

摘要

The planning target volume (PTV) concept has been created within the context of external beam radiotherapy (EBRT). It would be ideal to have a similar approach in brachytherapy (BT) to compensate for uncertainties. However, the BT and EBRT dose distributions are profoundly different, and the role of a PTV concept in BT needs a distinct discussion. The EBRT PTV concept is based on expanding the dose distribution into a homogeneous plateau reaching beyond the CTV. According to basic BT physics, there are significant dose gradients around radioactive source positions, and it is impossible to create homogeneous dose plateaus. This means that PTV margins cannot be directly applied in BT. Application of PTV margins in lateral and anterior-posterior directions can even lead to a significant and overall dose escalation ( approximately 8% per mm margin applied) for the individual patient and for the entire patient population. In the specific direction along the intrauterine tandem, safety margins can partly account for uncertainties, though. In conclusion, safety margins can only be partially applied in intracavitary BT, and it is not recommended to perform PTV delineation. The PTV seems not to be useful for dose reporting, and dose normalisation to PTV is strongly discouraged since it can lead to dose escalation.
机译:规划目标量(PTV)概念是在外部束放射疗法(EBRT)的背景下创建的。理想的是在近距离放射治疗(BT)中采用类似的方法来弥补不确定性。但是,BT和EBRT的剂量分布有很大不同,PTV概念在BT中的作用需要进行独特的讨论。 EBRT PTV概念基于将剂量分布扩展到CTV以外的均匀平台。根据基本的BT物理学,放射源位置附近存在明显的剂量梯度,并且不可能创建均匀的剂量平稳期。这意味着PTV余量不能直接应用于BT。在侧面和前后方向上应用PTV边缘甚至可能导致单个患者和整个患者群体发生显着的总体剂量提升(每mm边缘大约施加8%)。但是,沿着子宫内串联的特定方向,安全裕度可以部分解释不确定性。总之,安全裕度只能部分应用于腔内BT,不建议进行PTV划界。 PTV似乎不适用于剂量报告,并且强烈建议不要将PTV的剂量归一化,因为它可能导致剂量增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号