首页> 外文期刊>Radiotherapy and oncology: Journal of the European Society for Therapeutic Radiology and Oncology >ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilo-voltage cone-beam CT.
【24h】

ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilo-voltage cone-beam CT.

机译:ExacTrac X射线6自由度图像指导用于颅内无创立体定向放射治疗:与千伏锥束CT的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: To compare the residual setup errors measured with ExacTrac X-ray 6 degree-of-freedom (6D) and cone-beam computed tomography (CBCT) for a head phantom and patients receiving intracranial non-invasive fractionated stereotactic radiotherapy (SRT). MATERIALS AND METHODS: Setup data were collected on a Novalis Tx treatment unit for an anthropomorphic head phantom and 18 patients with intracranial tumors. Initial corrections were determined and corrected with the ExacTrac system only, and then the residual setup error was determined by means of three different procedures. These procedures included registrations of ExacTrac X-ray images with the corresponding digitally reconstructed radiographs (DRRs) using the ExacTrac 6D fusion, and registrations of CBCT images with the planning CT using both online 3D fusion and offline 6D fusion. The difference in residual setup errors between ExacTrac system and CBCT was computed. The impact of rotations on the difference was evaluated. RESULTS: A modest difference in residual setup errors was found between ExacTrac system and CBCT. The root-mean-square (RMS) of the differences observed for translations was typically <0.5mm for phantom, and <1.5mm for patients, respectively. The RMS of the differences for rotation(s) was however <0.2 degree for phantom, and <1.0 degree for patients, respectively. The impact of rotation on the setup difference was minor but not negligible. CONCLUSIONS: This study indicates that there is a general agreement between ExacTrac system and CBCT.
机译:背景与目的:比较用ExacTrac X射线6自由度(6D)和锥形束计算机断层扫描(CBCT)测量的头模和接受颅内无创分式立体定向放射治疗(SRT)的患者的残留设置误差)。材料与方法:在Novalis Tx治疗单元上收集了拟人化的头部幻影和18例颅内肿瘤患者的设置数据。仅使用ExacTrac系统确定并校正了最初的校正,然后通过三种不同的程序确定了残留设置误差。这些过程包括使用ExacTrac 6D融合将ExacTrac X射线图像与相应的数字重建射线照片(DRR)配准,以及使用在线3D融合和离线6D融合将CBCT图像与计划CT配准。计算了ExacTrac系统和CBCT之间残留设置误差的差异。评估了旋转对差异的影响。结果:在ExacTrac系统和CBCT之间发现了残留设置误差的适度差异。译文中观察到的差异的均方根(RMS)通常对于幻影分别<0.5mm,对于患者<1.5mm。但是,幻影旋转差异的RMS分别为<0.2度和患者<1.0度。旋转对设置差异的影响很小,但不可忽略。结论:这项研究表明ExacTrac系统和CBCT之间有一个普遍的共识。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号