The influence of geometrical changes on the dose distribution after I-125 seed implantation of the prostate.

机译:几何变化对I-125种子植入前列腺后剂量分布的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: After prostate implantation, dose calculation is usually based on a single imaging session, assuming no geometrical changes occur during the months of dose accumulation. In this study, the effect of changes in anatomy and implant geometry on the dose distribution was investigated. MATERIALS AND METHODS: One day, 1 month and 312 months after seed implantation, a combined TRUS-CT scan was made of 13 patients. Based on these scans changes in dose rate distribution were determined in prostate, urethra and bladder and a 'geometry corrected' dose distribution was estimated. RESULTS: When based on the day-1 scan, parameters representing high dose volumes in prostate and urethra were largely underestimated: V150 of the prostate 18+/-10% and V120 of the urethra 47+/-32%. The dose to a 2cm(3) hotspot in the bladder wall (D2cc), however, was overestimated by 31+/-35%. Parameters based on scans 1 month post-implant or later were all within +/-5% of geometry corrected values. CONCLUSION: Values meant to indicate the adequacy of dose coverage of the prostate, V100 and D90, were not influenced by geometrical changes and were independent of the post-implant scan date. Other parameters representing high dose volumes changed strongly within the first month after implantation.
机译:目的:前列腺植入后,剂量计算通常基于单个成像,假设剂量累积的几个月内没有发生几何变化。在这项研究中,研究了解剖结构和植入物几何形状的变化对剂量分布的影响。材料与方法:种子植入后第1天,1个月和312个月,对13例患者进行了TRUS-CT联合扫描。基于这些扫描,确定了前列腺,尿道和膀胱中剂量率分布的变化,并估算了“几何校正”剂量分布。结果:在第1天扫描的基础上,代表前列腺和尿道高剂量的参数被大大低估了:前列腺的V150为18 +/- 10%,尿道的V120为47 +/- 32%。但是,膀胱壁上2cm(3)热点的剂量(D2cc)被高估了31 +/- 35%。植入后1个月或更晚基于扫描的参数均在几何校正值的+/- 5%之内。结论:旨在表明前列腺,V100和D90剂量覆盖范围是否适当的值不受几何变化的影响,并且与植入后的扫描日期无关。代表高剂量体积的其他参数在植入后的第一个月内发生了很大变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号