An assessment of the use of skin flashes in helical tomotherapy using phantom and in-vivo dosimetry.

机译:使用幻影和体内剂量测定法评估螺旋闪光疗法中皮肤闪光的使用。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: In helical tomotherapy the nature of the optimizing and planning systems allows the delivery of dose on the skin using a build-up compensating technique (skin flash). However, positioning errors or changes in the patient's contour can influence the correct dosage in these regions. This work studies the behavior of skin-flash regions using phantom and in-vivo dosimetry. MATERIALS AND METHODS: The dosimetric accuracy of the tomotherapy planning system in skin-flash regions is checked using film and TLD on phantom. Positioning errors are induced and the effect on the skin dose is investigated. Further a volume decrease is simulated using bolus material and the results are compared. RESULTS: Results show that the tomotherapy planning system calculates dose on skin regions within 2 SD using TLD measurements. Film measurements show drops of dose of 2.8% and 26% for, respectively, a 5mm and 10mm mispositioning of the phantom towards air and a dose increase of 9% for a 5mm shift towardstissue. These measurements are confirmed by TLD measurements. A simulated volume reduction shows a similar behavior with a 2.6% and 19.4% drop in dose, measured with TLDs. CONCLUSION: The tomotherapy system allows adequate planning and delivery of dose using skin flashes. However, exact positioning is crucial to deliver the dose at the exact location.
机译:背景与目的:在螺旋断层扫描中,优化和计划系统的性质允许使用累积补偿技术(皮肤闪光)在皮肤上输送剂量。但是,定位误差或患者轮廓的变化会影响这些区域的正确剂量。这项工作使用幻影和体内剂量测定法研究皮肤闪光区域的行为。材料与方法:使用幻影上的胶片和TLD检查在皮肤毛刺区域的tomotherapy计划系统的剂量学准确性。引起定位误差并研究对皮肤剂量的影响。此外,使用大剂量材料模拟了体积的减少,并对结果进行了比较。结果:结果表明,Tomotherapy计划系统使用TLD测​​量来计算2 SD内皮肤区域的剂量。胶片测量结果表明,对于幻影朝向空气的5mm和10mm错位,剂量下降分别为2.8%和26%,向组织偏移5mm的剂量则增加了9%。这些测量通过TLD测量得到确认。模拟的体积减小显示出类似的行为,用TLD测​​量的剂量分别下降了2.6%和19.4%。结论:断层摄影系统允许使用皮肤闪光剂适当规划和分配剂量。然而,精确的定位对于在精确的位置输送剂量至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号