Observer variability when evaluating patient movement from electronic portal images of pelvic radiotherapy fields.

机译:从骨盆放射治疗领域的电子门禁图像评估患者的运动时,观察者的差异性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: A study has been performed to evaluate inter-observer variability when assessing pelvic patient movement using an electronic portal imaging device (EPID). MATERIALS AND METHODS: Four patient image sets were used with 3-6 portal images per set. The observer group consisted of nine radiographers with 3-18 months clinical EPID experience. The observers outlined bony landmarks on a digital simulator image and used matching software to evaluate field placement errors (FPEs) on each portal image relative to the reference simulator image. Data were evaluated statistically, using a two-component analysis of variance technique, to quantify both the inter-observer variability in evaluating FPEs and inter-fraction variability in patient position relative to the residuals of the analysis. Intra-observer variability was also estimated using four of the observers carrying out three sets of repeat readings. RESULTS: Eight sets of variance data were analysed, based on FPEs in two orthogonal directions for each of the four patient image sets studied. Initial analysis showed that both inter-observer variation and inter-fraction-patient position variation were statistically significant (P<0.05) in seven of the eight cases evaluated. The averaged root-mean-square (RMS) deviation of the observers from the group mean was 1.1mm, with a maximum deviation of 5.0mm recorded for an individual observer. After additional training and re-testing of two of the observers who recorded the largest deviations from the group mean, a subsequent analysis showed the inter-observer variability for the group to be significant in only three of the eight cases, with averaged RMS deviation reduced to 0.5mm, with a maximum deviation of 2.7mm. The intra-observer variability was 0.5mm, averaged over the four observers tested. CONCLUSIONS: We have developed a quantitative approach to evaluate inter-observer variability in terms of its statistical significance compared to inter-fraction patient movement. This will assist us in training and assessing observers required to perform this task on a routine basis.
机译:背景与目的:已经进行了一项研究,以评估使用电子门禁成像设备(EPID)评估骨盆患者运动时的观察者间差异。材料与方法:使用四个患者图像集,每组3-6个门静脉图像。观察员组由9位放射成像师组成,他们具有3-18个月的临床EPID经验。观察者在数字模拟器图像上勾勒出了骨标志,并使用匹配软件评估了相对于参考模拟器图像的每个门户图像上的场放置误差(FPE)。使用二元方差分析技术对数据进行统计评估,以量化评估FPE时观察者之间的变异性和患者位置相对于分析残差的部分间变异性。还使用四名观察员中的四名进行了三组重复读数来估计观察者内部的变异性。结果:针对四个研究图像集的每个,在两个正交方向上基于FPE分析了八组方差数据。初步分析显示,在评估的8例病例中,有7例观察者间差异和患者间分数差异均有统计学意义(P <0.05)。观察者与组平均值的平均均方根(RMS)偏差为1.1mm,单个观察者记录的最大偏差为5.0mm。在对两名记录的与组平均值偏差最大的观察员进行了额外的培训和重新测试之后,随后的分析表明,在八种情况中,只有三组的观察者间差异性显着,平均RMS偏差降低了到0.5mm,最大偏差为2.7mm。观察者内变异性为0.5mm,是所测试的四个观察者的平均值。结论:我们已经开发了一种定量的方法来评估观察者之间的差异,该差异是与小部分患者间运动相比的统计学意义。这将有助于我们培训和评估执行常规任务所需的观察员。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号