Tangential breast irradiation: a multi-centric intercomparison of dose using a mailed phantom and thermoluminescent dosimetry.

机译:切线乳房照射:使用邮寄的体模和热发光剂量法进行剂量的多中心比对。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To measure the accuracy of radiation therapy of the breast planned with 2D and 3D algorithms. MATERIALS AND METHODS: The accuracy of radiation therapy of the breast with 2D and 3D algorithms was investigated as a national intercomparison using a semi-anatomical breast phantom. The dose was measured with thermoluminescent (TL) dosemeters. RESULTS: The mean deviation of measured to planned dose at isocentre was -2.7%. The influence of some planning and irradiation factors is evaluated. CONCLUSION: The study demonstrates that beam energy and the use of CT have no marked influence on the accuracy of dose calculations, care has to be exercised when wedges are used and even with sophisticated 3D algorithms there is a systematic error in the dose received by the patient.
机译:目的:测量使用2D和3D算法计划的乳房放射治疗的准确性。材料与方法:使用半解剖乳房幻像作为全国比较研究了使用2D和3D算法对乳房进行放射治疗的准确性。用热发光(TL)剂量计测量剂量。结果:在等心点处与计划剂量的平均偏差为-2.7%。评估了一些计划和辐射因素的影响。结论:研究表明,束能量和CT的使用对剂量计算的准确性没有显着影响,使用楔形物时必须格外小心,即使采用复杂的3D算法,被检者所接收的剂量也存在系统误差。患者。