首页> 外文期刊>Radiotherapy and oncology: Journal of the European Society for Therapeutic Radiology and Oncology >Daily positioning accuracy of frameless stereotactic radiation therapy with a fusion of computed tomography and linear accelerator (focal) unit: evaluation of z-axis with a z-marker.
【24h】

Daily positioning accuracy of frameless stereotactic radiation therapy with a fusion of computed tomography and linear accelerator (focal) unit: evaluation of z-axis with a z-marker.

机译:计算机断层扫描和线性加速器(焦距)单元的融合,无框架立体定向放射疗法的每日定位精度:使用z标记评估z轴。

获取原文
获取原文并翻译 | 示例
           

摘要

To evaluate quantitative positioning errors of frameless stereotactic radiation therapy with a fusion of computed tomography (CT) and linear accelerator unit, Z-type CT markers were attached to patients, and CT images were obtained before and after daily treatment. In 40 verification tests, geometrical errors were never more than 1 mm.
机译:为了评估计算机断层扫描(CT)和线性加速器单元融合进行的无框立体定向放射治疗的定量定位误差,将Z型CT标记贴在患者身上,并在每日治疗前后获得CT图像。在40个验证测试中,几何误差从未超过1毫米。