The evaluation of an inexpensive, 2D, video based gait assessment system for clinical use

机译:用于临床的廉价,基于二维视频的步态评估系统的评估

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to investigate the clinical potential of an augmented-video-based-portable-system (AVPS). The AVPS included a walkway grid mat made of vinyl flooring, flat paper bull's eye markers, four photoswitches mounted on tripods, a light-indicator, a video camera, and a computer with ProTrainer System software. The AVPS output was compared to a "gold standard" 3D Vicon Motion Analysis System both statically and dynamically over a fixed range (-90° to +90°) using a two-segment-goniometric-rig marked with both bull's eye and retroreflective markers. At each segment angle position, three trials of data were captured. The reliability of the AVPS was also tested using three raters. Further twelve, young, healthy subjects participated in a concurrent validity study in which they performed six gait trials which were simultaneously recorded by both systems. Both motion analysis systems showed low levels of intra subject variability in all kinematic variables indicated by the size of the standard deviations across the six trials. There were no significant differences between the motion systems with respect to the kinematic variables (P > 0.05). The results showed a high intra- and inter-rater reliability for both the kinematic and temporo-spatial parameters. With respect to gait events the lowest ICC value for the intra-rater reliability test was 0.993 for the kinematic variables, and ranged from 0.941 to 0.956 for the temporo-spatial variables and 0.731 to 0.954 for the tibia inclination angles. The validation data suggest the AVPS is capable of generating highly reliable and repeatable data when applied to normal subjects and could be used within the clinical setting.
机译:这项研究的目的是调查基于增强视频的便携式系统(AVPS)的临床潜力。 AVPS包括一个由乙烯基地板制成的走道网格垫,平面纸牛眼标记,安装在三脚架上的四个光电开关,一个指示灯,一台摄像机以及一台装有ProTrainer System软件的计算机。使用标有牛眼和回射标记的两段式测角仪,在固定范围(-90°至+ 90°)上将静态和动态地将AVPS输出与“黄金标准” 3D Vicon运动分析系统进行了比较。 。在每个节段角位置,捕获了三个试验数据。还使用三个评估者测试了AVPS的可靠性。另有十二名健康的年轻受试者参加了一项并行效度研究,其中他们进行了六个步态试验,两个系统同时记录了该步态试验。两种运动分析系统在所有运动学变量中均显示出较低的受试者内部变异性,这由六项试验中的标准偏差的大小表示。运动系统之间在运动学变量方面无显着差异(P> 0.05)。结果表明,运动学和时空参数均具有较高的评价者内部和评价者之间的可靠性。关于步态事件,评估者内可靠性测试的最低ICC值在运动学变量上为0.993,在时间空间变量上为0.941至0.956,在胫骨倾斜角上为0.731至0.954。验证数据表明,AVPS在应用于正常受试者时能够生成高度可靠和可重复的数据,并且可以在临床环境中使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号