首页> 外文期刊>Interactive cardiovascular and thoracic surgery >Off-bypass implantation of a self-expandable valved stent between inferior vena cava and right atrium1^
【24h】

Off-bypass implantation of a self-expandable valved stent between inferior vena cava and right atrium1^

机译:在下腔静脉和右心房之间进行旁道植入自膨胀瓣式支架1 ^

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A glutaraldehyde preserved valved bovine jugular xenograft mounted in a nitinol 'Z' stent, expandable from 7 to 28 mm of internal diameter, was evaluated in vitro (column of water developing a pressure of 45 mmHg and a mock loop including a pulsatile pump) and in vivo in five adult pigs with intra-vascular ultrasound to measure the inferior vena cava diameter via a retroperitoneal access. Through a stent-graft delivery system (24 French) the self expandable valved stent was implanted off-bypass in the inferior vena cava, between hepatic veins and cavo-atrial junction, with flow and pressure gradient recording. The mean length of the valved stent was 22.80(+-) 1.06 mm, the mean internal diameter 20.97(+-) 0.5 mm and the mean external diameter 26.67(+-) 0.9 mm. The valve leaking under pressure was 32.5(+-) 12.3 ml/min. The mean pressure gradient recorded across the self expandable valved stent implanted in the inferior vena cava was 1.0(+-) 0.5 mmHg (range 0-2 mmHg). Intra-vascular ultrasound showed partial opening and closing of the valve (mean area reduction from 148.5 to 81.5 mm~2), with almost complete occlusion only during deep breaths. The in vitro and in vivo experiments confirmed the feasibility of potential application of the self-expandable valved stent implanted off-bypass in the inferior vena cava for late conversion of failing total cavo-pulmonary connection; intra-vascular ultrasound allows for adequate implantation and evaluation.
机译:体外评估了戊二醛保存的带瓣膜牛颈异种移植物,该移植物安装在镍钛合金“ Z”型支架上,可从内径从7毫米扩展到28毫米(水柱的压力为45毫米汞柱,模拟回路包括脉动泵),并且在五只成年猪体内进行血管内超声检查,以通过腹膜后通路测量下腔静脉直径。通过一个支架移植物输送系统(24 French),将可自我扩张的带瓣膜支架以旁路方式植入肝静脉和腔-房结之间的下腔静脉,并记录流量和压力梯度。带瓣膜支架的平均长度为22.80(±)1.06 mm,平均内径为20.97(±)0.5 mm,平均外径为26.67(±)0.9 mm。阀在压力下泄漏为32.5(±)12.3 ml / min。记录在下腔静脉内植入的自扩张瓣膜支架上的平均压力梯度为1.0(±)0.5 mmHg(范围为0-2 mmHg)。血管内超声显示瓣膜部分打开和关闭(平均面积从148.5减小至81.5 mm〜2),仅在深呼吸时几乎完全闭塞。体外和体内实验证实了将可扩展的带瓣膜支架旁路植入下腔静脉潜在应用的可行性,该方法可用于晚期失败的全腔肺连接。血管内超声可以进行适当的植入和评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号