首页> 外文期刊>Interactive cardiovascular and thoracic surgery >Fracture and cardiac migration of an implanted venous catheter
【24h】

Fracture and cardiac migration of an implanted venous catheter

机译:植入式静脉导管的骨折和心脏迁移

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Distal embolization of a fractured indwelling central catheter is a rare life-threatening complication. Interventional radiologic techniques are often successfully used to retrieve the fractured tip but in rare cases, a surgical intervention is needed. We present a case in which radiologic methods failed to remove the embolized fragment because of multiple venous occlusions, and a cardiotomy was finally necessary.After right atriotomy, the fractured embolized tip was successfully removed, as well as the proximal portion. Implanted catheters should be removed after completion of treatment, or the integrity of the system should be monitored on a regular basis.
机译:骨折的留置中心导管远端栓塞是一种罕见的危及生命的并发症。介入放射学技术通常可以成功地用于恢复骨折的尖端,但在极少数情况下,需要进行手术干预。我们提出了一种情况,其中放射学方法由于多个静脉阻塞而无法去除栓塞碎片,最后需要进行切开术。右房室切除术后,成功地去除了破裂的栓塞尖端以及近端部分。治疗结束后应移除植入的导管,或者应定期监测系统的完整性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号