Nitric oxide breath testing by tunable-diode laser absorption spectroscopy: application in monitoring respiratory inflammation

机译:可调谐二极管激光吸收光谱对一氧化氮的呼气测试:在监测呼吸道炎症中的应用

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We used a high-resolution mid-IR tunable-laser absorption spectroscopy (TLAS) system with a single IV-VI laser operating near 5.2 μm to measure the level of exhaled nitric oxide (eNO) in human breath. A method of internal calibration using simultaneous eNO and exhaled CO_(2) measurements eliminated the need for system calibration with gas standards. The results observed from internally calibrating the instrument for eNO measurements were compared with measurements of eNO calibrated to gas standards and were found to be similar. Various parameters of the TLAS system for eNO breath testing were examined and include gas cell pressure, exhalation time, and ambient NO concentrations. A reduction in eNO from elevated concentrations (~44 parts in 10~(9)) to near-normal levels (<20 parts in 10~(9)) from an asthmatic patient was observed after the patient had received treatment with an inhaled glucocorticoid anti-inflammatory medication. Such measurements can help in evaluating airway inflammation and in monitoring the effectiveness of anti-inflammatory therapies.
机译:我们使用高分辨率中红外可调谐激光吸收光谱(TLAS)系统,并在5.2μm附近操作单个IV-VI激光,以测量人呼吸中呼出的一氧化氮(eNO)的水平。使用同时进行eNO和呼出CO_(2)测量的内部校准方法消除了使用气体标准进行系统校准的需要。将内部校准仪器进行eNO测量所观察到的结果与根据气体标准校准的eNO测量结果进行了比较,发现结果相似。检查了用于eNO呼气测试的TLAS系统的各种参数,包括气室压力,呼气时间和环境NO浓度。在患者接受吸入糖皮质激素治疗后,观察到哮喘患者的eNO从浓度升高(在10〜(9)中约44份)降至接近正常水平(在10〜(9)中<20份)。抗炎药。这样的测量可以帮助评估气道炎症并监测抗炎疗法的有效性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号