Using the National Institute of Standards and Technology high-accuracy cryogenic radiometer (HACR), we have realized a scale of absolute spectral response between 406 and 920 nm. The HACR, an electrical-substitution radiometer operating at cryogenic temperatures, achieves a combined relative standard uncertainty of 0.021%. Silicon photodiode light-trapping detectors were calibrated against the HACR with a typical relative standard uncertainty of 0.03% at nine laser wavelengths between 406 and 920 nm. Modeling of the quantum efficiency of these detectors yields their responsivity throughout this range with comparable accuracy. [References: 29]
机译:使用美国国家标准技术研究院的高精度低温辐射计(HACR),我们已经实现了406至920 nm之间的绝对光谱响应范围。 HACR是一种在低温下运行的电子替代辐射计,可实现0.021%的组合相对标准不确定度。相对于HACR校准了硅光电二极管光阱检测器,在406至920 nm之间的九个激光波长下,典型相对标准不确定度为0.03%。这些探测器的量子效率建模可在整个范围内以相当的精度产生其响应度。 [参考:29]