A family of sensors is presented with fluorescence decay-time measurements used as the sensing technique. The concept is to take a single fluorophore with a suitably long fluorescence decay time as the basic building block for numerous different sensors. Analyte recognition can be performed by different functional groups that are necessary for selective interaction with the analyte. To achieve this, the principle of excited-state electron transfer is applied with pyrene as the fluorophore. Therefore the same instrumentation based on a small, ambient air-nitrogen laser and solid-state electronics can be used to measure different analytes, for example, oxygen, pH, carbon dioxide, potassium, ammonium,lead, cadmium, zinc, and phosphate. (C) 1996 Optical Society of America [References: 35]
机译:提出了一系列传感器,其中荧光衰减时间测量用作传感技术。概念是将具有适当较长荧光衰减时间的单个荧光团作为众多不同传感器的基本构件。可以通过与分析物进行选择性相互作用所必需的不同官能团来执行分析物识别。为了实现这一点,采用pyr作为荧光团来应用激发态电子转移的原理。因此,基于小型环境空气氮气激光器和固态电子设备的同一仪器可用于测量不同的分析物,例如氧气,pH,二氧化碳,钾,铵,铅,镉,锌和磷酸盐。 (C)1996年美国眼镜学会[参考:35]