首页> 外文期刊>Applied optics >SeaWiFS transfer-to-orbit experiment
【24h】

SeaWiFS transfer-to-orbit experiment

机译:SeaWiFS移轨实验

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We present the results of an experiment designed to measure the changes in the radiometric calibration of the Sea-viewing Wide Field-of-view Sensor (SeaWiFS) from the time of its manufacture to the time of the start of on-orbit operations. The experiment uses measurements of the Sun at the manufacturer's facility to predict the instrument outputs during solar measurements immediately after launch. Because an onboard diffuser plate is required for these measurements, the experiment measures changes in the instrument-diffuser system. There is no mechanism in this experiment to separate changes in the diffuser from changes in the instrument. For the eight SeaWiFS bands, the initial instrument outputs on orbit averaged 0.8% higher than predicted with a standard deviation of 0.9%. The greatest difference was 2.1% (actual output higher than predicted) for band 3. The estimated uncertainty for the experiment is 3%. Thus the transfer-to-orbit experiment shows no changes in the radiometric sensitivities of the SeaWiFS bands—at the 3% level—from the completion of the instrument's manufacture to its insertion into orbit.
机译:我们提供了一个实验结果,该实验旨在测量从制造到开始在轨运行期间的海景宽视场传感器(SeaWiFS)的辐射校准的变化。该实验使用制造商工厂的太阳测量值来预测发射后立即进行的太阳测量期间的仪器输出。由于这些测量需要使用板载扩散板,因此实验可测量仪器-扩散器系统中的变化。此实验中没有机制可将扩散器的变化与仪器的变化区分开。对于八个SeaWiFS波段,初始仪器在轨输出平均比预期高0.8%,标准偏差为0.9%。频段3的最大差异为2.1%(实际输出高于预期)。实验的估计不确定度为3%。因此,从轨道转移实验表明,从完成仪器制造到将其插入轨道,SeaWiFS波段的辐射敏感性(在3%的水平上)都没有变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号