首页> 外文期刊>Applied optics >Method for characterization of filter radiometers
【24h】

Method for characterization of filter radiometers

机译:滤波辐射计的表征方法

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We have developed a new method for characterizing the irradiance responsivity of filter radiometers. The method is based on a spatially uniform, known irradiance, generated by combining several identical laser beams. The measurement setup and the experimental demonstration at one wavelength are presented. The diffraction correction related to the generated irradiance is studied experimentally. The uncertainty analysis of the method indicates a relative staudard uncertainty of 1 × 10~(-3). The results with the new method are compared with the characterization measurements based on our present spectral-irradiance scale. The results have a relative deviation of 1 × 10(-3), which is well within the combined standard uncertainty of the comparison.
机译:我们已经开发出一种表征滤光辐射计辐射响应度的新方法。该方法基于通过组合几个相同的激光束而产生的空间均匀的已知辐照度。介绍了一种波长下的测量设置和实验演示。实验研究了与产生的辐照度有关的衍射校正。该方法的不确定度分析表明,标准斯托达德不确定度为1×10〜(-3)。将新方法的结果与基于我们目前的光谱辐照度标度的表征测量结果进行比较。结果的相对偏差为1×10(-3),完全在比较的组合标准不确定度之内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号