The O-arm Surgical Imaging System (Medtronic Sofamor Danek, Memphis, TN, USA) is an increasingly popular tool for spinal surgical procedures that allows for intraoperative acquisition of multi-dimensional spinal imaging that can be used either to confirm placement of spinal instrumentation or to guide spinal screw placement in conjunction with a navigation system. The machine is typically placed open from the side of the patient positioned prone on the Jackson spinal table (Mizuho OSI, Union City, CA, USA) and then closed to complete a ring around the patient to acquire images. A failure of the O-arm opening mechanism can lead to a patient becoming effectively trapped within the device with limited physician access to the patient's body, a situation that may be remedied by using a manual opening procedure. This report highlights a patient safety issue encountered when a mechanical failure of the O-arm occurred and one of the two tools required for its manual opening was missing. We describe the improvised method employed to successfully open the device.
机译:O形臂外科成像系统(Medtronic Sofamor Danek,Memphis,Tn,USA)是脊柱外科手术的越来越流行的工具,允许术中采集多维脊椎成像,可以使用以确认脊柱仪器的放置或引导与导航系统结合的脊柱螺钉放置。该机器通常从患者侧面放置,位于杰克逊脊柱表(Mizuho Osi,Union City,Ca,USA),然后关闭以完成患者的环以获取图像。 O形臂开口机构的故障可能导致患者有效地捕获在设备内的设备内,该器件有限地访问患者的身体,这种情况可以通过使用手动打开过程来补救。此报告突出显示O形臂的机械故障时遇到的患者安全问题,并且缺少其手动开口所需的两种工具之一。我们描述了用于成功打开设备的简易方法。