Voluntary patient motion is a common cause of image degradation during MRI and leads to repeated scanning, decreasing efficiency, and increasing costs. We hypothesized that providing an educational pamphlet to patients before their MRI examination could improve image quality and decrease the number of repeated sequences needed because of motion artifacts.Over 12 months, we recruited patients undergoing MRI for any neurologic condition. The control group received a routine safety questionnaire concerning MRI scanning. The intervention group was given an additional pamphlet describing the examination and graphically emphasizing the value of remaining still during scanning; comprehension was confirmed by questionnaire. The radiology technologists performing the examinations were blinded to group assignments; they recorded the number of repeated sequences needed because of motion artifacts and assessed image quality on a scale of 0 to 4 (0 = unusable, 4 = perfect).The number of patients requiring repeated MRI sequences (control group vs intervention group: 40 vs 20, respectively; p = 0.009) and the total number of repeated MRI sequences (52 vs 27, p = 0.004) decreased in the group who read the pamphlet compared with the control group.Providing a simple educational pamphlet to patients before their MRI examinations that illustrated motion degradation and emphasized the need to remain still significantly reduced the number of repeated sequences deemed necessary by the MRI technologist.
机译:患者的自愿运动是MRI期间图像质量下降的常见原因,并导致重复扫描,效率降低和成本增加。我们假设在进行MRI检查之前为患者提供教育手册可以改善图像质量并减少由于运动伪影而导致的重复序列的数量。在过去的12个月中,我们招募了接受过MRI检查的任何神经系统疾病患者。对照组收到有关MRI扫描的常规安全调查表。干预组得到了另外一本小册子,描述了检查内容,并以图形方式强调了扫描过程中静止状态的价值。通过问卷调查确认了理解。进行检查的放射技师对小组任务不知情。他们记录了由于运动伪影而需要的重复序列的数量,并以0到4的等级评估了图像质量(0 =无法使用,4 =完美)。需要重复MRI序列的患者数量(对照组vs干预组:40 vs与对照组相比,阅读该手册的组分别减少了20,p = 0.009)和重复MRI序列的总数(52 vs 27,p = 0.004)。为患者提供了一个简单的教育手册该图说明了运动的退化,并强调了保持显着降低MRI技术人员认为必要的重复序列数量的必要性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号