Partial Arc Beam Filtration: A Novel Approach to Reducing CT Breast Radiation Dose.

机译:部分弧束过滤:减少CT乳腺辐射剂量的新颖方法。

获取原文
获取原文并翻译 | 示例
           

摘要

We sought to assess the effectiveness of a novel CT radiation dose reduction strategy in which filtration was added at the x-ray tube output port between the x-ray beam and the breast area of three sizes of anthropomorphic phantoms.To evaluate the dose-reduction potential of partial arc x-ray beam filtration, copper foil filtration or lead foil filtration was placed over CT scanners' covers when scanning anthropomorphic phantoms representative of a 5-year-old child, a 10-year-old child, and an adult female. Dose reduction was calculated as the percentage difference between the mean entrance radiation dose (on the phantoms' surfaces at locations representing the sternum and left breast) in unshielded scans compared with the mean dose in scans shielded by copper or lead foil. We also compared the CT numbers and noise sampled in regions representing the lung and the soft tissues near the sternum, left breast, and spine in CT images of the phantoms during unshielded scans relative to acquisitions shielded by copper or lead foil.Entrance dose at the sternum and left breast in the three anthropomorphic phantoms was reduced by 28-66% and 54-79% when using copper or lead foil filtration, respectively. However, copper foil filtration affected the CT numbers and noise in the CT images less than the lead foil filtration did (8.2% vs 32% mean increase in noise).By incorporating partial arc beam filtration into CT scanners, substantial dose reductions may be achieved with a minimal increase in image noise.
机译:我们试图评估一种新颖的CT辐射剂量减少策略的有效性,该策略是在三种尺寸的拟人幻影的X射线束与乳房区域之间的X射线管输出端口处添加过滤。在扫描代表5岁儿童,10岁儿童和成年女性的拟人化体模时,将局部电弧X射线束过滤,铜箔过滤或铅箔过滤的潜力置于CT扫描仪的封面上。减少剂量的计算方法是:在非屏蔽扫描中,平均入射辐射剂量(代表胸骨和左乳房的体模表面上的平均辐射剂量)与由铜或铅箔屏蔽的扫描中的平均剂量之间的百分比差异。我们还比较了在非屏蔽扫描期间相对于被铜或铅箔屏蔽的采集,在代表幻影的CT图像中代表肺和胸骨,左乳房和脊柱附近的软组织区域的CT数和噪声进行了比较。使用铜或铅箔过滤时,三种拟人化体模中的胸骨和左乳房分别减少了28-66%和54-79%。然而,铜箔过滤对CT数量的影响和CT图像中的噪声比铅箔过滤的影响要小(8.2%对比32%的平均噪声增加)。通过将部分电弧束过滤纳入CT扫描仪中,可以显着降低剂量图像噪声的增加最小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号