首页> 外文期刊>AJR: American Journal of Roentgenology : Including Diagnostic Radiology, Radiation Oncology, Nuclear Medicine, Ultrasonography and Related Basic Sciences >CT screening and follow-up of lung nodules: effects of tube current-time setting and nodule size and density on detectability and of tube current-time setting on apparent size.
【24h】

CT screening and follow-up of lung nodules: effects of tube current-time setting and nodule size and density on detectability and of tube current-time setting on apparent size.

机译:CT筛查和肺结节的随访:管电流时间设置,结节大小和密度对可检测性的影响以及管电流时间设置对表观大小的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: The purpose of the study was to quantify and compare the effect of CT dose and of size and density of nodules on the detectability of lung nodules and to quantify the influence of CT dose on the size of the nodules. MATERIALS AND METHODS: From 50 patients a total of 125 cuboidal regions of interest (3 x 3 x 1.5 cm volumes) showing a single nodule (
机译:目的:本研究的目的是量化和比较CT剂量以及结节大小和密度对肺结节可检测性的影响,并量化CT剂量对结节大小的影响。材料与方法:从50例患者中,总共选择了125个长方体感兴趣的区域(3 x 3 x 1.5厘米体积),显示了单个结节(≤8 mm)和27个正常长方体。使用软件从模拟不同剂量水平的原始数据重建图像集:300(原始剂量),220、180、140、100、80、60、50、40、30、20、10和5个参考mAs。使用逻辑回归模型分析三个盲读者的可检测性。对于小于3 mm的结节尺寸和大于3 mm的结节尺寸,以及-300 HU的结节衰减以及大于-300 HU的结节衰减,计算比值。结果:与300 mAs的标准设置相比,10 mAs或更大的电子管电流时间设置与单个读取器灵敏度的显着差异无关。在5 mAs下,只有一个读取器的灵敏度从82%显着降低到77%(p = 0.0035)。根据比值比和逻辑回归结果,可以假设低结节密度,小结节大小和剂量水平对灵敏度的最强负面影响。 5至300 mAs之间的平均结节体积测量误差为2.2%+/- 18%(SD),远低于观察者间体积测量误差率38%+/- 45%。结论:结果显示了低剂量CT方案在10 mAs下进行肺结节随访的可行性。肺结节随访中的计算机辅助体积测量可减少观察者之间的差异。