Variation in tube voltage for adult neck MDCT: Effect on radiation dose and image quality

机译:成人颈部MDCT的管电压变化:对辐射剂量和图像质量的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE. The purpose of this study was to assess the effect of peak kilovoltage on radiation dose and image quality in adult neck MDCT. MATERIALS AND METHODS. An anthropomorphic phantom with metal oxide semiconductor field effect transistor detectors was imaged with a 64-MDCT scanner. The reference CT protocol called for 120 kVp, and images obtained with that protocol were compared with CT images obtained with protocols entailing 80, 100, and 140 kVp. All imaging was performed with automatic tube current modulation. Organ dose and effective dose were determined for each protocol and compared with those obtained with the 120-kVp protocol. Image noise was evaluated objectively and subjectively for each protocol. RESULTS. The highest organ doses for all protocols were to the thyroid, ocular lens, skin, and mandible. The greatest reductions in organ dose were for the bone marrow of the cervical spine and mandible: 43% and 35% with the 100-kVp protocol and 63% and 53% with the 80-kVp protocol. Effective dose decreased as much as 9% with the 100-kVp protocol and 12% with the 80-kVp protocol. Use of the 140-kVp protocol was associated with an increase in organ dose as high as 64% for bone marrow in the cervical spine and a 19% increase in effective dose. Image noise increased with lower peak kilovoltage. The measured noise difference was greatest at 80 kVp, absolute increases were less than 2.5 HU. There was no difference in subjective image quality among protocols. CONCLUSION. Reducing the voltage from 120 to 80 kVp for neck CT can result in greater than 50% reduction in the absorbed organ dose to the bone marrow of the cervical spine and mandible without impairment in subjective image quality.
机译:目的。这项研究的目的是评估峰值千伏电压对成人颈部MDCT的辐射剂量和图像质量的影响。材料和方法。用64-MDCT扫描仪对带有金属氧化物半导体场效应晶体管检测器的拟人模型进行成像。参考CT协议要求120 kVp,并将通过该协议获得的图像与通过80、100和140 kVp协议获得的CT图像进行比较。所有成像均通过自动管电流调制进行。确定每种方案的器官剂量和有效剂量,并将其与120-kVp方案获得的器官剂量和有效剂量进行比较。针对每种方案客观地和主观地评估图像噪声。结果。所有方案的最高器官剂量为甲状腺,眼镜,皮肤和下颌骨。器官剂量的最大减少是颈椎和下颌骨的骨髓:100 kVp方案分别为43%和35%,80 kVp方案则分别为63%和53%。使用100-kVp协议时,有效剂量降低了9%,而使用80-kVp协议时降低了12%。 140 kVp方案的使用与颈椎骨髓器官剂量增加高达64%和有效剂量增加19%有关。图像噪声随峰值千伏电压的降低而增加。在80 kVp处测得的噪声差异最大,绝对增量小于2.5 HU。协议之间的主观图像质量没有差异。结论。将用于颈部CT的电压从120 kVp降低到80 kVp可以导致吸收到颈椎和下颌骨的骨髓的器官剂量减少50%以上,而不会损害主观图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号