Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies.

机译:患者居中不当的CT自动管电流调制系统的误操作:幻像研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: Inappropriate patient centering on the gantry changes the size of the localizer radiographs used for CT examinations, influencing the operation of CT automatic tube current modulation because tube current is controlled with information from localizer radiographs. The purpose of this study was to examine the influence of inappropriate patient centering on the gantry isocenter on automatic tube current modulation. MATERIALS AND METHODS: An elliptical phantom was scanned with four automatic tube current modulation techniques after acquisition of localizer radiographs in the horizontal and vertical directions with the phantom center shifted from the gantry isocenter in the vertical direction. After scanning, the magnification rate of the frontal localizer radiographs, tube current-time product, and image noise were examined. RESULTS: On phantom studies, the magnification rate of localizer radiographs showed a linear relation to the vertical deviation of the phantom from the gantry isocenter. From 50 mm above to 50 mm below the gantry isocenter, tube current-time products ranged from 75% to 141% compared with those at the gantry isocenter. In addition, increases and decreases in the amount of image noise related to changes in tube current-time product were confirmed. CONCLUSION: Inappropriate patient centering causes misoperation of automatic tube current modulation systems, in which tube current is controlled with information from localizer radiographs, and thus causes increases in tube current or image noise.
机译:目的:以机架为中心的不适当患者会改变用于CT检查的定位器X射线照片的尺寸,从而影响CT自动电子管电流调制的操作,因为电子管电流受定位器X射线照片的信息控制。这项研究的目的是检查不适当的患者居中在龙门架等中心线上对自动管电流调制的影响。材料与方法:在获取水平和垂直方向的定位射线照片后,采用四种自动管电流调制技术扫描椭圆体模,其中体模中心在垂直方向上偏离龙门等中心线。扫描后,检查额叶定位器射线照片的放大率,管电流时间乘积和图像噪声。结果:在体模研究中,定位器射线照片的放大率与体模与门架等中心线的垂直偏差呈线性关系。从龙门等中心上方的50 mm到龙门等中心以下的50 mm,与龙门等中心的管电流时间乘积的范围为75%至141%。另外,证实了与管电流-时间乘积的变化有关的图像噪声量的增加和减少。结论:不正确的患者对中会导致自动电子管电流调制系统误操作,在自动电子管电流调制系统中,电子管电流由定位器射线照片中的信息控制,从而导致电子管电流或图像噪声增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号