3-T navigator parallel-imaging coronary MR angiography: targeted-volume versus whole-heart acquisition.

机译:3-T导航仪平行成像冠状动脉MR血管造影:目标体积与全心采集。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: The purpose of this study was to compare whole-heart acquisition with targeted-volume acquisition in 3-T navigator coronary MR angiography with parallel imaging. SUBJECTS AND METHODS: The right and left coronary arteries of 20 subjects were imaged with axial whole-heart acquisition and two oblique targeted-volume acquisitions. RESULTS: Both whole-heart and targeted-volume acquisitions were completed with similar navigator efficiencies ( approximately 50%) and depicted similar coronary artery diameters ( approximately 3 mm) (p >or= 0.06). The lengths of the coronary arteries were not significantly different (p = 0.07-0.45) for the whole-heart and targeted-volume approaches. Depiction of the sharper coronary arteries (p
机译:目的:本研究的目的是比较3-T导航仪冠状动脉MR血管造影与并行成像中的全心脏采集与目标体积采集。研究对象和方法:采用轴向全心脏采集和两次倾斜目标体积采集对20名受试者的左右冠状动脉成像。结果:全心和目标体积的采集均以相似的导航效率(约50%)完成,并描绘出相似的冠状动脉直径(约3 mm)(p>或= 0.06)。对于全心脏和目标体积方法,冠状动脉的长度没有显着差异(p = 0.07-0.45)。使用目标体积方法可更好地描绘更尖的冠状动脉(p <或= 0.04)和整体图像质量(p <0.02)。结论:对于当前的3T导航仪平行成像冠状动脉MR血管造影,与全心脏采集相比,目标体积采集可产生更清晰的冠状动脉图像。