Gas bubble motion artifact in MDCT.

机译:MDCT中的气泡运动伪影。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: The purpose of our study was to characterize the imaging features of an MDCT artifact caused by gas bubble motion. MATERIALS AND METHODS: For the period 2002-2006, the CT images of 10 patients that revealed a curvilinear artifact thought to be due to a gas bubble moving during CT acquisition were retrospectively identified and reviewed by an attending body radiologist. The clinical images were acquired on MDCT scanners. A phantom containing water and moving air bubbles (gas injection rates of 0.1 and 0.5 mL/s) was designed, built, and scanned using a 16-MDCT scanner. Computer simulation was used to test our hypothesis that these observed artifacts originated from moving gas bubbles. RESULTS: Semicircular tubular CT artifacts with attenuation close to that of air were observed in all 10 clinical cases. The acquired MDCT images from the phantom and the computer simulations showed tubular air-attenuation semicircular artifacts that closely matched the imaging findings on the 10 clinical cases. Increased rates of bubble injection multiplied the artifact. Based on the computer simulation, the precise appearance of the artifact depends on the bubble location and velocity relative to the rotation of the CT scanner. CONCLUSION: Gas bubble motion during CT generates a semicircular air-attenuation artifact that has not been previously described to our knowledge. The artifact is typically found in bowel and other liquid-containing structures in which a bubble of gas floats up through a liquid medium during CT.
机译:目的:我们的研究目的是表征由气泡运动引起的MDCT伪影的成像特征。材料与方法:在2002年至2006年期间,由一名身体放射科医生对10例患者的CT图像进行了回顾性鉴定,并对其进行了回顾,这些患者发现了曲线伪影,这些伪影被认为是由于在CT采集过程中出现了气泡移动所致。临床图像在MDCT扫描仪上获取。设计,制造并使用16-MDCT扫描仪设计了包含水和运动气泡(注入速度分别为0.1和0.5 mL / s)的体模。使用计算机仿真来检验我们的假设,即这些观察到的伪影源自移动的气泡。结果:在所有10例临床病例中均观察到衰减程度接近空气衰减的半圆形管状CT伪影。从体模获取的MDCT图像和计算机模拟显示出管状的空气衰减半圆形伪影,这些伪影与10例临床病例的影像学发现非常匹配。气泡注入速率的增加使伪影倍增。基于计算机模拟,伪影的精确外观取决于气泡位置和相对于CT扫描仪旋转的速度。结论:CT期间的气泡运动会产生半圆形的空气衰减伪影,这是我们以前所没有描述的。该伪影通常出现在肠和其他含液体的结构中,在CT扫描过程中,气泡会通过液体介质向上漂浮。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号