首页> 外文期刊>Medical dosimetry: official journal of the American Association of Medical Dosimetrists >Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit
【24h】

Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit

机译:来自不同系统伽马射线的Gafchromic EBT3薄膜校准分析:伽马刀和钴-60单位

获取原文
获取原文并翻译 | 示例
           

摘要

In recent years, Gafchromic films are used as an advanced instrument for dosimetry systems. The EBT3 films are a new generation of Gafchromic films. Our main interest is to compare the response of the EBT3 films exposed to gamma rays provided by the Theratron 780C as a conventional radiotherapy system and the Leksell Gamma Knife as a stereotactic radiotherapy system (SRS). Both systems use Cobalt-60 sources, thus using the same energy. However, other factors such as source-to-axis distance, number of sources, dose rate, direction of irradiation, shape of phantom, the field shape of radiation, and different scatter contribution may influence the calibration curve. Calibration curves for the 2 systems were measured and plotted for doses ranging from 0 to 40 Gy at the red and green channels. The best fitting curve was obtained with the Levenberg-Marquardt algorithm. Also, the component of dose uncertainty was obtained for any calibration curve. With the best fitting curve for the EBT3 films, we can use the calibration curve to measure the absolute dose in radiation therapy. Although there is a small deviation between the 2 curves, the p-value at any channel shows no significant difference between the 2 calibration curves. Therefore, the calibration curve for each system can be the same because of minor differences. The results show that with the best fitting curve from measured data, while considering the measurement uncertainties related to them, the EBT3 calibration curve can be used to measure the unknown dose both in SRS and in conventional radiotherapy. (C) 2017 Published by Elsevier Inc.
机译:近年来,Gafchromic薄膜用作剂量测定系统的先进仪器。 EBT3薄膜是新一代Gafchromic薄膜。我们的主要兴趣是将暴露于Theratron 780C提供的伽马射线的EBT3薄膜作为常规放射治疗系统和Leksell Gamma刀作为立体定向放射治疗系统(SRS)进行比较。两种系统都使用钴-60来源,从而使用相同的能量。然而,诸如源到轴距离的其他因素,源源,剂量率,照射方向,辐射形状,辐射的场形状,以及不同的散射贡献可能影响校准曲线。测量2个系统的校准曲线并绘制用于在红色和绿色通道的0至40 gy的剂量。用Levenberg-Marquardt算法获得了最佳的拟合曲线。而且,为任何校准曲线获得剂量不确定性的组分。对于EBT3薄膜的最佳拟合曲线,我们可以使用校准曲线来测量放射疗法中的绝对剂量。虽然2曲线之间存在小的偏差,但是在任何通道处的p值显示出2校准曲线之间没有显着差异。因此,由于微小的差异,每个系统的校准曲线可以是相同的。结果表明,通过测量数据的最佳拟合曲线,在考虑与它们相关的测量不确定性的同时,EBT3校准曲线可用于测量SRS和常规放射疗法中的未知剂量。 (c)2017年由elsevier公司发布

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号