Development and pilot testing of a new electromyography device - EMG Lavita

机译:新型电拍摄装置的开发和试验试验 - EMG Lavita

获取原文
获取原文并翻译 | 示例
           

摘要

The mechanisms involved in muscle adaptations from different resistance training protocols are still not completely understood. The aim of this study is to develop an EMG device with a supervisory system that records intensity and time duration of concentric and eccentric phases of a muscular contraction. An electronic circuit was developed with an instrumentation amplifier INA118P, a bandpass filter 30 a 150 Hz, and a rectifier. The analog signal was converted to a digital signal using a 10-bit A/D converter that is part of the Arduino NANO (Atmega328p) microcontroller, which sends data via serial communication to a supervisory system that processes and displays data in graphs and tables. A usability system test was performed by four experts using the SUS scale. Also, the system validation test was performed by the researcher who developed the device by recording simultaneously EMG signals from his muscle fibers using two different electromyography devices - EMG Comercial and EMG Lavita. For this test, the researcher performed five data collections of ten repetitions of the elbow flexion movement with a 1-minute interval for each collection. A Pearson's correlation coefficient analysis was performed with the RMS values previously collected from the devices (r=0.86), which indicated a good to excellent correlation. The device developed in this study had a similar behavior than the commercial and validated tool; besides, all experts evaluated the system as satisfactory.
机译:仍未完全理解来自不同电阻训练协议的肌肉改编的机制。本研究的目的是开发具有监控系统的EMG装置,其记录肌肉收缩的同心和偏心阶段的强度和时间持续时间。用仪表放大器INA118P,带通滤波器30a150Hz和整流器开发电子电路。使用10位A / D转换器将模拟信号转换为数字信号,该转换器是Arduino Nano(Atmega328P)微控制器的一部分,其通过串行通信将数据发送到处理和显示图形和表中的数据。使用SUS规模的四个专家执行可用性系统测试。此外,系统验证测试由研究人员执行,该研究人员通过使用两种不同的肌电纤维使用两种不同的电拍摄设备来记录从他的肌肉纤维的同时的EMG信号进行。对于该测试,研究人员对每个收集的1分钟间隔进行了五次重复的肘部屈曲运动的五个数据收集。使用先前从器件(R = 0.86)收集的RMS值进行Pearson的相关系数分析,该RMS值表示良好的良好相关性。本研究开发的设备具有比商业和验证工具类似的行为;此外,所有专家均以令人满意的评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号