Influence of different sized files on the accuracy of two electronic apex locators

机译:不同大小文件对两个电子APEX定位器精度的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract The aim of this study was to evaluate the impact of different file sizes on the accuracy of two electronic apex locators ( EAL s). Thirty extracted human single‐rooted permanent mandibular incisors were used. A #10 K‐file was inserted in the root canal until its end could be observed (using a light microscope) through the apical foramen. One millimetre was subtracted to establish working length ( WL ). Electronic readings were performed using MiniApex Locator or Root ZX II , from #10 K‐file to #130 K‐file. Statistical analysis was performed by two‐way anova and Tukey test ( P ≤ 0.05). From #60 to #130 K‐file, observed differences were noted between the values obtained with both EAL s and WL ( P ≤ 0.05). The MiniApex Locator showed increased means when measurements were made with #50 to #70 and with #120 ( P = 0.008) and #130 ( P = 0.005) K‐files. File sizes influenced the accuracy of EAL s – the greater the instrumentation size, the higher mean differences compared to WL .
机译:摘要本研究的目的是评估不同文件大小对两个电子顶尖定位器(EAL S)的准确性的影响。使用三十次提取的人单生根永久性下颌切牙。在根运河中插入#10 k文件,直到可以通过顶端孔观察(使用光学显微镜)的端部。减去一毫米以建立工作长度(WL)。使用MiniaPex定位器或根ZX II进行电子读数,从#10 k文件到#130 k文件。通过双向ANOVA和Tukey检验进行统计分析(P≤0.05)。从#60到#130 k文件,在使用EAL S和WL的值之间注意到观察到的差异(P≤0.05)。当用#50到#70和#120(p = 0.008)和#130(p = 0.005)k文件进行测量时,MiniaPex定位器显示出增加的意味着。文件大小影响了EAL S的准确性 - 仪器尺寸越大,与WL相比的平均差异越高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号