首页> 外文期刊>Applied optics >Calibration-free self-absorption model for measuring nitric oxide concentration in a pulsed corona discharge
【24h】

Calibration-free self-absorption model for measuring nitric oxide concentration in a pulsed corona discharge

机译:用于测量脉冲电晕放电中一氧化氮浓度的免校准自吸收模型

获取原文
获取原文并翻译 | 示例
           

摘要

The effect of self-absorption on emission intensity distributions can be used for species concentration measurements. A calculation model is developed based on the Beer-Lambert law to quantify this effect. And then, a calibration-free measurement method is proposed on the basis of this model by establishing the relationship between gas concentration and absorption strength. The effect of collision parameters and rotational temperature on the method is also discussed. The proposed method is verified by investigating the nitric oxide emission bands (A~2 Σ~+ → X~2 Π) that are generated by a pulsed corona discharge at various gas concentrations. Experiment results coincide well with the expectations, thus confirming the precision and accuracy of the proposed measurement method.
机译:自吸收对发射强度分布的影响可用于物种浓度测量。基于比尔-兰伯特定律开发了计算模型以量化此影响。然后,通过建立气体浓度与吸收强度之间的关系,在该模型的基础上提出了无标定测量方法。还讨论了碰撞参数和旋转温度对方法的影响。通过研究在各种气体浓度下脉冲电晕放电产生的一氧化氮发射带(A〜2Σ〜+→X〜2Π),验证了该方法的有效性。实验结果与预期吻合得很好,从而证实了所提出的测量方法的准确性和准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号