We analyze two important points related to the experimental emissivity measurements in this paper: the radiometer calibration accuracy and its stability to determine the required frequency of calibration. The usual two-temperature calibration method is compared to a more accurate method, which uses the measurement of blackbody radiation at several temperatures. Additionally, the suitability of the two-temperature method is studied as a function of the gap between both temperatures. Differences higher than 200 deg C are needed to obtain an acceptable calibration. The temporal stability of the calibration and the influence of the environmental conditions are also analyzed.
机译:在本文中,我们分析了与实验发射率测量相关的两个要点:辐射计校准精度及其稳定性,以确定所需的校准频率。将通常的两温校准方法与更精确的方法进行比较,后者使用了在多个温度下黑体辐射的测量值。此外,根据两种温度之间的差异研究了两种温度方法的适用性。需要高于200摄氏度的差异以获得可接受的校准。还分析了校准的时间稳定性以及环境条件的影响。