首页> 外文期刊>Applied optics >Flow-through integrating cavity absorption meter: experimental results
【24h】

Flow-through integrating cavity absorption meter: experimental results

机译:流通式集成腔吸收仪:实验结果

获取原文
获取原文并翻译 | 示例
           

摘要

We report experimental results from a flow-through integrating cavity absorption meter. The operating range of the device is from 0.004 m~(-1) to over 80m~(-1) of absorption. Absorption coefficients have been measured with 8percent or less change in the presence of over 200 m~(-1) of scattering in the medium. The instrument signal has been shown to be independent of flow rate up to 20 liters/min and thus independent of turbulence. This large operational range along with the ability to measure absorption independently of adverse scattering affects allows the instrument to be utilized in a wide range of environmental conditions.
机译:我们报告了流经集成腔吸收仪的实验结果。该设备的工作范围是从0.004 m〜(-1)到超过80m〜(-1)的吸收。在介质中存在超过200 m〜(-1)的散射的情况下,吸收系数的变化幅度为8%或更少。仪器信号已显示出与最高20升/分钟的流速无关,因此与湍流无关。如此大的操作范围以及独立于不利散射影响而测量吸收的能力,使该仪器可在广泛的环境条件下使用。