首页> 外文期刊>Applied optics >Improved method for calibrating a Stokes polarimeter
【24h】

Improved method for calibrating a Stokes polarimeter

机译:校准斯托克斯旋光仪的改进方法

获取原文
获取原文并翻译 | 示例
           

摘要

We present a method for calibrating a polarization state analyzer that uses a set of well-characterized reference polarization states and makes no assumptions about the optics contained in the polarimeter other than their linearity. The method requires that a matrix be constructed that contains the data acquired for each of the reference polarization states and that this matrix be pseudoinverted. Since this matrix is usually singular, we improve the method by performing the pseudoinversion by singular value decomposition, keeping only the four largest singular values. We demonstrate the calibration technique using an imaging polarimeter based upon liquid crystal variable retarders and with light emitting diode (LED) illumination centered at 472 nm, 525 nm, and 630 nm. We generate the reference polarization states by using an unpolarized source, a single polarizer, and a Fresnel rhomb. This method is particularly useful when calibrations are performed on field-grade instruments at a centrally maintained facility and when a traceability chain needs to be maintained.
机译:我们提出了一种用于校准偏振态分析仪的方法,该方法使用了一组性能良好的参考偏振态,除了线性度之外,没有对偏振计中包含的光学器件做出任何假设。该方法要求构造一个矩阵,该矩阵包含为每个参考偏振态获取的数据,并且该矩阵是伪反转的。由于此矩阵通常是奇异的,因此我们通过奇异值分解执行伪反转来改进方法,仅保留四个最大的奇异值。我们演示了使用基于液晶可变延迟器的成像偏振计和以472 nm,525 nm和630 nm为中心的发光二极管(LED)照明进行校准的技术。我们通过使用非偏振光源,单个偏振器和菲涅耳菱形来生成参考偏振态。当在集中维护的场所对现场级仪器执行校准时,以及需要维护可追溯性链时,此方法特别有用。