首页> 外文期刊>Applied optics >Implementation of a loss-compensated recirculating delayed self-heterodyne interferometer for ultranarrow laser linewidth measurement
【24h】

Implementation of a loss-compensated recirculating delayed self-heterodyne interferometer for ultranarrow laser linewidth measurement

机译:用于超窄激光线宽测量的损耗补偿循环延迟自外差干涉仪的实现

获取原文
获取原文并翻译 | 示例
           

摘要

Ultranarrow laser linewidth measurement using an optimized loss-compensated recirculating delayed self-heterodyne interferometer is described. An experimental setup is constructed to measure subkilohertz laser linewidths. The system parameters are optimized to obtain the best beat signals. The experimental results agree well with the theoretical analysis. Two methods of linewidth interpretation are presented and analyzed based on the experimental results. It is proved that a loss-compensated recirculating delayed self-heterodyne interferometer is an effective tool for measuring an ultranarrow laser linewidth.
机译:描述了使用优化的损耗补偿再循环延迟自外差式干涉仪进行的超窄激光线宽测量。构建了一个实验装置来测量亚千赫兹激光的线宽。优化系统参数以获得最佳拍频信号。实验结果与理论分析吻合良好。根据实验结果,提出和分析了两种线宽解释方法。事实证明,损耗补偿的循环延迟自外差干涉仪是测量超窄激光线宽的有效工具。